Abstract :
The effect of methanol, acetone and acetonitrile on the sensitivity, selectivity and the detection limits (LODs) of the determination of chromium species by ion chromatography was investigated. A collinear dual-beam thermal lens spectrometer was used for the direct detection of chromium complexes [pre-column derivatized Cr(III)–pyridine-2,6-dicarboxylic acid, and post-column derivatized Cr(VI)–1,5-diphenylcarbazide] following the ion chromatographic separation on a Dionex HPIC-CS5A solvent compatible column. Different amounts of organic solvents were added directly to the eluent (up to 30%) and to the post-column reagent (up to 60%) to improve the thermooptical properties of the solvents. Consequently, the sensitivity of the technique was increased by a factor of 2–3 and LODs of 0.1 and 10 μg dm−3 were achieved for Cr(VI) and Cr(III), respectively, when the eluent reaching the detector contained 30% of acetonitrile. The addition of organic solvents also resulted in significant changes in retention times, which improved the Cr(III)/Cr(VI) separation.