Title of article :
Impact of an error in the column hold-up time for correct adsorption isotherm determination in chromatography: II. Can a wrong column porosity lead to a correct prediction of overloaded elution profiles?
Author/Authors :
Samuelsson، نويسنده , , Jِrgen and Zang، نويسنده , , Jia and Murunga، نويسنده , , Anne and Fornstedt، نويسنده , , Torgny and Sajonz، نويسنده , , Peter، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
8
From page :
205
To page :
212
Abstract :
The adsorption isotherm was determined for phenol in methanol/water on a C-8 stationary phase using frontal analysis in staircase mode, assuming different total column porosities, from 1 to 87%. Each set of adsorption isotherm data, with a certain column porosity, was fitted to various adsorption models and the generated parameters were used to calculate overloaded elution band profiles that were compared with experiments. It was found that the bi-Langmuir model had an optimum fit for a porosity that corresponds well with the value found experimentally. The adsorption energy distribution (AED) calculations and error analysis confirmed a bimodal energy distribution. It was also found that band profiles can be accurately predicted with a quite arbitrary chosen porosity, under prerequisite that a wrong but flexible adsorption model is chosen instead of the correct one. The latter result is very useful for quick optimizations of preparative separations where the exact value of the column porosity is not available.
Keywords :
Hold-up volume , void volume , Equilibrium isotherm , Frontal analysis , Adsorption energy distribution , Ideal model , Prediction of band profiles , Column porosity
Journal title :
Journal of Chromatography A
Serial Year :
2008
Journal title :
Journal of Chromatography A
Record number :
1510933
Link To Document :
بازگشت