• Title of article

    Low electroosmotic flow measurement by tilting microchip

  • Author/Authors

    Zhou، نويسنده , , Fang and Wang، نويسنده , , Wei and Wu، نويسنده , , Wen-Ya and Zhang، نويسنده , , Jianrong and Zhu، نويسنده , , Jun-Jie، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    4
  • From page
    221
  • To page
    224
  • Abstract
    A novel method for low electroosmotic flow (EOF) rates measurement by tilting microchip which based upon the hydrostatic pressure conception and sampling zone method is described. Sampling zone could be detected in the tilting microchip but not in non-tilting one due to the hydrostatic pressure driven. The method is fulfilled to calculate low EOF rates by detecting the liquid flow velocity driven by hydrostatic pressure, and difference between the apparent mobility of the migrating analyte in two modes is caused by the effect of hydrostatic pressure. And then the EOF rates in unknown low EOF microchip can be calculated. Different microchannels modified with bovine serum albumin (BSA), myoglobin (MB) and polyvinyl alcohol (PVA) were used to verify the method, the EOF rate value was 1.73 ± 0.03, 1.21 ± 0.05, 0.34 ± 0.04 × 10−4 cm2 V−1 s−1, respectively. The results obtained by the proposed method were agreed well with conventional methods.
  • Keywords
    Low EOF measurement , hydrostatic pressure , Tilting microchip , Modified microchannels
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography A
  • Record number