Title of article :
Development of packed-column suppressor system for capillary ion chromatography and its application to environmental waters
Author/Authors :
Sedyohutomo، Anang نويسنده , , Anang and Lim، نويسنده , , Lee Wah and Takeuchi، نويسنده , , Toyohide، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
A novel suppressor unit for capillary ion chromatography was designed to reduce the background conductivity and at the same time to increase the analyte signal. Regeneration of the suppressor was carried out on-line by passing an appropriate acidic solution through the column to displace the accumulated eluent cations. By using two 6-port microswitching valves and two packed capillary column suppressors, the background conductivity of sodium carbonate–bicarbonate mobile phase was maintained at low conductivity for continuous chromatographic runs, and the detection limits at low ppb levels were achieved. The relative standard deviations (RSDs) for the retention time, peak area and peak height of six common inorganic anions (0.05 mM each of F−, Cl−, NO2−, Br−, NO3− and SO42−) were between 0.5–0.9, 1.1–4.6 and 0.7–4.9%, respectively. The present system was successfully applied to the determination of inorganic anions in river water and tap water samples.
Keywords :
Common inorganic anions , Packed capillary column suppressor , Environmental water analysis , Continuous suppression , Capillary ion chromatography , 6-Port microswitching valve
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A