Title of article
Effect of capillary cross-section geometry and size on the separation of proteins in gradient mode using monolithic poly(butyl methacrylate-co-ethylene dimethacrylate) columns
Author/Authors
Marc Nischang، نويسنده , , Ivo and Svec، نويسنده , , Frantisek and Fréchet، نويسنده , , Jean M.J.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
7
From page
2355
To page
2361
Abstract
Porous polymer monoliths have been prepared in capillaries with circular or square cross-sections and lateral dimensions of 50, 75, 100 μm as well as in a rectangular 38 μm × 95 μm capillary. These capillaries have been used to determine the effect of the size and shape of their cross-section on the porous and hydrodynamic properties of poly(butyl methacrylate-co-ethylene dimethacrylate) monoliths. The capillaries were studied by scanning electron microscopy and evaluated for their permeability to flow and their performance in the liquid chromatographic separation of a protein mixture comprising ribonuclease A, cytochrome c, myoglobin, and ovalbumin using a linear gradient of acetonitrile in the mobile phase. No differences resulting from channel geometry were found for the various capillary columns. These results demonstrate that standard capillaries with circular geometry are a good and affordable alternative conduit for modeling the processes carried out in microfluidic chips with a variety of geometries.
Keywords
scalability , Porous polymer monolith , Nano-LC , Conduit shape , Proteins , Reversed phase HPLC
Journal title
Journal of Chromatography A
Serial Year
2009
Journal title
Journal of Chromatography A
Record number
1511759
Link To Document