Title of article :
Simultaneous determination of diastereoisomeric and enantiomeric impurities in SSS-octahydroindole-2-carboxylic acid by chiral high-performance liquid chromatography with pre-column derivatization
Author/Authors :
Wang، نويسنده , , Jin Zhao and Zeng، نويسنده , , Su and Hu، نويسنده , , Gong Yun and Wang، نويسنده , , Dan Hua، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
4
From page :
3044
To page :
3047
Abstract :
SSS-Octahydroindole-2-carboxylic acid (SSS-Oic) is a key intermediate used in the synthesis of some angiotensin-converting enzyme (ACE) inhibitors. The separation of diastereoisomers and enantiomers of Oic was performed using a pre-column derivatization chiral HPLC method. Phenyl isothiocyanate (PITC) was used as the derivatization reagent. Three PITC derivatives of Oic stereoisomers were separated on an Ultron ES-OVM chiral column (150 mm × 4.6 mm, 5 μm). Derivatization conditions such as reaction temperature, reaction time and derivatization reagent concentration were investigated. The chromatographic conditions for separation of the three PITC-Oic derivatives were optimized. The method was successfully applied in the diastereoisomeric and enantiomeric purity test of SSS-Oic.
Keywords :
Diastereoisomers , Octahydroindole-2-carboxylic acid , Enantiomers , Pre-column derivatization