• Title of article

    Measurement of the dose deposition characteristics of x-ray fluoroscopy beams in water

  • Author/Authors

    Alaei، Parham نويسنده , , Gerbi، Bruce J. نويسنده , , Geise، Richard A. نويسنده , , Fetterly، Kenneth A. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    -204
  • From page
    205
  • To page
    0
  • Abstract
    The purpose of this study was to characterize the x-ray dose distribution of fluoroscopy beams by measuring their percent depth dose curves and lateral dose profiles in a water phantom. Percent depth dose curves were measured near the surface with an Attix parallel plate chamber and deep within the water phantom with a Farmer-type cylindrical chamber. Percent depth dose curves were compared to published data where applicable. Lateral dose profiles were measured at depths of 2, 5, 10, and 15 cm in phantom with a Farmer chamber. Pulsed, 50 mA x-ray beams with peak tube potentials of 60, 80, 100, and 120 kV and half value layers of 1.89, 2.52, 3.20, and 4.09 mm Al, respectively, were investigated.
  • Keywords
    Fault current limiter , short circuit current , transient over voltage , power quality
  • Journal title
    MEDICAL PHYSICS
  • Serial Year
    2001
  • Journal title
    MEDICAL PHYSICS
  • Record number

    1512