Title of article :
New 18-l process-scale counter-current chromatography centrifuge
Author/Authors :
Sutherland، نويسنده , , Ian and Hewitson، نويسنده , , Peter and Ignatova، نويسنده , , Svetlana، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
5
From page :
4201
To page :
4205
Abstract :
A new Dynamic Extractions Maxi-counter-current chromatography (CCC) centrifuge with a column volume of 18-l has been installed in the Advanced Bioprocessing Centre at Brunel. This instrument has four times the capacity of the 4.6-l Maxi-CCC centrifuge which has been operating robustly for 3 years. Tests using the model sample system benzyl alcohol and p-cresol with a heptane:ethyl acetate:methanol:water (HEMWat) phase system (1.4:0.1:0.5:1.0) show that resolution is almost double with this new high capacity device. Commissioning tests with a mixture of caffeine, KD = 0.21; ferulic acid, KD = 0.82; umbelliferone, KD = 1.2 and vanillin, KD = 1.49 using a HEMWat phase system of 1:1.5:1:1.5 on the 9-l column show that resolutions equivalent to analytical instruments will be possible using the full 18-l capacity. They also show that predictable scale-up from simple test tube tests is feasible with knowledge of the stationary phase retention for the planned process scale run.
Keywords :
Process scale , Counter-current chromatography , CCC , guess