Title of article
Examination of the temporal effect in a flow injection analysis system using multi-channel absorbance detection
Author/Authors
Lin، نويسنده , , Shu-Hui and Chang، نويسنده , , Kai-Chieh and Chen، نويسنده , , Yu-Chun and Yu، نويسنده , , Tiing and Pai، نويسنده , , Su-Cheng، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
7
From page
4618
To page
4624
Abstract
In addition to the Poiseuille effect, a so-called temporal effect was proposed recently to elucidate the commonly observed tailing peak signals of flow injection analysis (FIA). A multi-channel absorbance detector was used in this study to obtain the FIA peaks on both the spatial and temporal coordinates. The temporal effect was analyzed by comparison of the profiles between the experimental and the corresponding Gaussian peaks, and by comparison of asymmetry factors between the spatial and the temporal peaks. The temporal effect appeared to be the major factor under flow rates ranging from 0.5 to 8 ml min−1. This was despite the presence of a spatially frontal peak observed in the FIA tubing, which was found to result in a tailing peak on the temporal coordinate due to this discussed cause. In addition, the temporal effect became greater as the flow rate increased.
Keywords
Poiseuille effect , Temporal effect , Flow injection analysis , Asymmetry factor , Tailing peaks , Multi-channel detection