Title of article :
Determination of phenylalanine in blood by high-performance anion-exchange chromatography–pulsed amperometric detection to diagnose phenylketonuria
Author/Authors :
Jeong، نويسنده , , Ji-Seon and Sim، نويسنده , , Hee-Jung and Lee، نويسنده , , Yong-Moon and Yoon، نويسنده , , Hye-Ran and Lee، نويسنده , , Dong Hwan and Hong، نويسنده , , Seon-Pyo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
6
From page :
5709
To page :
5714
Abstract :
We have developed a high-performance anion-exchange chromatography with pulsed amperometric detection method for the detection of phenylalanine (Phe) and diagnosis of phenylketonuria (PKU). Sample pretreatment steps were simplified without derivatization. The analyte was separated within 5 min. The detection limit (S/N = 3) for Phe was 50 pg. Linear dynamic range was 1.23–14.43 mg/dL (r2 = 0.9999) for a dried blood spot. The mean recoveries of Phe for intra- and inter-day assays were found to be 96.87–104.16%. This method clearly differentiated PKU-positive groups from normal groups, and proved to be a practical procedure for rapid screening and follow-up monitoring of PKU.
Keywords :
Phenylketonuria , Dried blood spot , phenylalanine , Rapidity , high throughput , Pulsed amperometric detection
Journal title :
Journal of Chromatography A
Serial Year :
2009
Journal title :
Journal of Chromatography A
Record number :
1512218
Link To Document :
بازگشت