Title of article :
Carbon nanotube field-effect transistor detector associated to gas chromatography for speciation of benzene, toluene, ethylbenzene, (o-, m- and p-)xylene
Author/Authors :
Silva، نويسنده , , Lurdes I.B. and Ferreira، نويسنده , , Filipe D.P. and Rocha-Santos، نويسنده , , Teresa A.P. and Duarte، نويسنده , , A.C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
5
From page :
6517
To page :
6521
Abstract :
An analytical methodology based on a field-effect transistor detector using carbon nanotubes (NTFET) coupled to a gas chromatograph has been developed for the speciation of the following aromatic compounds: benzene, toluene, ethylbenzene, m-xylene, p-xylene and o-xylene (BTEX). This methodology combines the proven separation capability of gas chromatography (GC) with the potential for detection of a NTFET. The developed analyzer shows a high and stable analytical response upon repeated analysis of BTEX during 4 weeks, with detection limit less than 4 μg/L. The GC–NTFET system also shows a great suitability for actual monitoring of indoor atmospheres and no significant difference was observed between the results obtained by the developed analyzer and a more classical analytical methodology, namely gas chromatography–flame ionization detection (GC–FID).
Keywords :
Carbon nanotubes , BTEX detection , Field-effect transistor , Nanodetector