Title of article :
Rapid high performance liquid chromatography method development with high prediction accuracy, using 5 cm long narrow bore columns packed with sub-2 μm particles and Design Space computer modeling
Author/Authors :
Fekete، نويسنده , , Szabolcs and Fekete، نويسنده , , Jen? and Moln?r، نويسنده , , Imre and Ganzler، نويسنده , , Katalin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
Many different strategies of reversed phase high performance liquid chromatographic (RP-HPLC) method development are used today. This paper describes a strategy for the systematic development of ultrahigh-pressure liquid chromatographic (UHPLC or UPLC) methods using 5 cm × 2.1 mm columns packed with sub-2 μm particles and computer simulation (DryLab® package). Data for the accuracy of computer modeling in the Design Space under ultrahigh-pressure conditions are reported. An acceptable accuracy for these predictions of the computer models is presented. This work illustrates a method development strategy, focusing on time reduction up to a factor 3–5, compared to the conventional HPLC method development and exhibits parts of the Design Space elaboration as requested by the FDA and ICH Q8R1. Furthermore this paper demonstrates the accuracy of retention time prediction at elevated pressure (enhanced flow-rate) and shows that the computer-assisted simulation can be applied with sufficient precision for UHPLC applications (p > 400 bar). Examples of fast and effective method development in pharmaceutical analysis, both for gradient and isocratic separations are presented.
Keywords :
Computer-assisted optimization , Method development , UHPLC , Sub-2 , ?m particles , DryLab®
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A