Title of article :
Simultaneous determination of eleven quinolones in animal feed by liquid chromatography with fluorescence and ultraviolet absorbance detection
Author/Authors :
Galarini، نويسنده , , Roberta and Fioroni، نويسنده , , Laura and Angelucci، نويسنده , , Federico and Tovo، نويسنده , , Gloria R. and Cristofani، نويسنده , , Elisa، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
A rapid and simple multi-residue procedure is described for assaying eleven quinolones (cinoxacin, ciprofloxacin, danofloxacin, difloxacin, enrofloxacin, flumequine, marbofloxacin, nalidixic acid, norfloxacin, oxolinic acid and sarafloxacin) in feeds at sub-additive levels (1–5 mg kg−1). Five grams of sample were extracted by a metaphosphoric acid/acetonitrile mixture (70/30, v/v) and purified onto OASIS HLB cartridges. The determination was achieved by liquid chromatography (LC) using a GEMINI C18 analytical column both with fluorescence detection (FD) and photodiode-array (DAD). Limits of detection for each drug were in the range 0.04–0.8 mg kg−1. Above the limit of quantification (LOQ), in poultry feed the recoveries were from 69 to 98% with relative standard deviations less than or equal 10%. Finally the measurement uncertainty was estimated using the bottom-up approach.
Keywords :
Quinolones , feed , Liquid chromatography , Validation , uncertainty
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A