Title of article :
Evaluation of a condensation nucleation light scattering detector for the analysis of synthetic polymer by supercritical fluid chromatography
Author/Authors :
Takahashi، نويسنده , , Kayori and Kinugasa، نويسنده , , Shinichi and Yoshihara، نويسنده , , Ryo and Nakanishi، نويسنده , , Akira and Mosing، نويسنده , , Renee K. and Takahashi، نويسنده , , Rheo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
A condensation nucleation light scattering detector (CNLSD) was adapted for use as a detector for supercritical fluid chromatography. The performance of the CNLSD was evaluated and compared to evaporative light-scattering detector (ELSD) using a well-defined equimass mixture of uniform poly(ethylene glycol) oligomers and a certified reference material of poly(ethylene glycol) 1000. The CNLSD was able to detect a 10 times less concentrated solution of uniform oligomers compared to the ELSD. The quantitativeness of CNLSD was high enough to obtain the molecular mass distribution of poly(ethylene glycol) 1000 without any calibrations; on the other hand, the original data measured by ELSD was about 4% smaller than the certified value of poly(ethylene glycol) 1000. The CNLSD was suitable for supercritical fluid chromatography as a mass concentration detector.
Keywords :
ELSD , CNLSD , Calibration , Poly(ethylene glycol) , SFC , NQAD
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A