Title of article :
A protocol for the measurement of all the parameters of the mass transfer kinetics in columns used in liquid chromatography
Author/Authors :
Gritti، نويسنده , , Fabrice and Guiochon، نويسنده , , Georges، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
15
From page :
5137
To page :
5151
Abstract :
Band broadening in chromatography results from the combination of the dispersive effects that are associated with the different steps involved in the migration of compound bands along the column. These steps include longitudinal diffusion, trans-particle mass transfer, external film mass transfer, overall eddy diffusion, including trans-column, short-range inter-channel, trans-channel eddy diffusion, and the possible, additional mass transfer contributions arising from heat friction and the thermal heterogeneity of the column. We describe a series of experiments that provide the data needed to determine the coefficients of the contributions to band broadening of each one of these individual mass transfer steps. This specifically designed protocol can provide key information regarding the kinetic performance of columns used in liquid chromatography and explain why different columns behave so differently. The limitations, accuracy and precision of these methods are discussed. Further avenues of research that could improve the characterization of the mass transfer mechanisms in chromatographic columns, possibly contributing to the development of better columns, are suggested.
Keywords :
core-shell particles , Mass transfer , Liquid chromatography , Longitudinal diffusion , Eddy diffusion , External film mass transfer coefficient , Trans-particle mass transfer coefficient , Frictional heating , Totally porous particles
Journal title :
Journal of Chromatography A
Serial Year :
2010
Journal title :
Journal of Chromatography A
Record number :
1513274
Link To Document :
بازگشت