Title of article :
Simultaneous determination of polysorbate 20 and unbound polyethylene-glycol in protein solutions using new core–shell reversed phase column and condensation nucleation light scattering detection
Author/Authors :
Fekete، نويسنده , , Szabolcs and Ganzler، نويسنده , , Katalin and Fekete، نويسنده , , Jen?، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
A novel fast and sensitive method has been developed for the specific simultaneous determination of polysorbate 20 (Tween 20) and unbound polyethylene-glycol (PEG) from liquid formulations in the presence of proteins and excipients. The quantitative determination is based on a fast liquid chromatographic (HPLC) separation and condensation nucleation light scattering detection (CNLSD or NQAD™). The method uses a Kinetex core–shell column (100 mm × 3 mm, 2.6 μm) and methanol–water–trifluoroacetic acid mobile phase. The rapid HPLC-CNLSD method presented here is suitable for quantifying polysorbate 20 in the range of 10–60 μg/ml and unbound PEG in the range of 2–40 μg/ml in protein solutions within good manufacturing practices (GMP) of the pharmaceutical industry.
Keywords :
Fast chromatography , Tween 20 , PEG , Kinetex , NQAD , CNLSD
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A