Title of article
Improvement in the generation of adsorption isotherm data in the elution by characteristic points method—The ECP-slope approach
Author/Authors
Samuelsson، نويسنده , , Jِrgen and Undin، نويسنده , , Torgny and Tِrncrona، نويسنده , , Anders and Fornstedt، نويسنده , , Torgny، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
7
From page
7215
To page
7221
Abstract
The elution by characteristic points (ECP) method is a very rapid and precise method for determination of the phase system equilibrium of phase systems in broad solute concentration ranges. Thus, the method is especially suitable for rapid characterization of high efficient separation systems. One important source of error, the effects by the post-loop dispersion, was eliminated in a recent investigation. In this study, the systematic error caused by the selection of the integration starting point at concentration equal to 0 is eliminated. This is done by developing and validating a new procedure for isotherm data generation; the ECP-slope method. The method generates raw slope data of the adsorption isotherm instead of raw adsorption data by integrations as the classical ECP does. Both numerical and experimental data were used for the comparison of the classical ECP approach with the slope-ECP method.
Keywords
Single component adsorption , Adsorption isotherms , Elution by characteristic points method , The ECP-slope method , Slope data
Journal title
Journal of Chromatography A
Serial Year
2010
Journal title
Journal of Chromatography A
Record number
1513528
Link To Document