• Title of article

    A microfabricated graphitic carbon column for high performance liquid chromatography

  • Author/Authors

    Barrow، نويسنده , , D.A. and Castell، نويسنده , , O.K. and Sykes، نويسنده , , N. and Myers، نويسنده , , P. C. Ritchie، نويسنده , , H.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    5
  • From page
    1983
  • To page
    1987
  • Abstract
    We report the first development of a novel, planar, microfluidic, graphitic carbon separations column utilizing an array of graphitic micropillars of diamond cross-section as the chromatographic stationary phase. 795 nm femtosecond laser ablation was employed to subtractively machine fluidic architectures and a micropillared array in a planar, graphitic substrate as a monolithic structure. A sample injector was integrated on-chip, together with fluid-flow distribution architectures to minimize band-broadening and ensure sample equi-distribution across the micro-pillared column width. The separations chip was interfaced directly to the ESI probe of a Thermofisher Surveyor mass spectrometer, enabling the detection of test-mixture analytes following their differential retention on the micro-pillared graphitic column, thus demonstrating the exciting potential of this novel separations format. Importantly, unlike porous, graphitic microspheres, the temperature and pressure resilience of the microfluidic device potentially enables use in subcritical H2O chromatography.
  • Keywords
    Micropillars , graphitic carbon , CHIP , chromatography , Microfluidics , Femtosecond laser , HPLC
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2011
  • Journal title
    Journal of Chromatography A
  • Record number

    1513896