Title of article :
High efficiency, high temperature separations on silica based monolithic columns
Author/Authors :
Rogeberg، نويسنده , , Magnus and Wilson، نويسنده , , Steven Ray and Malerod، نويسنده , , Helle and Lundanes، نويسنده , , Elsa and Tanaka، نويسنده , , Nobuo and Greibrokk، نويسنده , , Tyge، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
8
From page :
7281
To page :
7288
Abstract :
The effect of temperature on separation using reversed-phase monolithic columns has been investigated using a nano-LC pumping system for gradient separation of tryptic peptides with MS detection. A goal of this study was to find optimal conditions for high-speed separations. The chromatographic performance of the columns was evaluated by peak capacity and peak capacity per time unit. Column lengths ranging from 20 to 100 cm and intermediate gradient times from 10 to 30 min were investigated to assess the potential of these columns in a final step separation, e.g. after fractionation or specific sample preparation. Flow rates from 250 to 2000 nL/min and temperatures from 20 to 120 °C were investigated. Temperature had a significant effect on fast separations, and a flow rate of 2000 nL/min and a temperature of 80 °C gave the highest peak capacity per time unit. These settings produced 70% more protein identifications in a biological sample compared to a conventional packed column. Alternatively, an equal amount of protein identifications was obtained with a 40% reduction in run time compared to the conventional packed column.
Keywords :
Protein identification , Peptide separation , high temperature , Peak capacity , Fast separation , Silica monolith
Journal title :
Journal of Chromatography A
Serial Year :
2011
Journal title :
Journal of Chromatography A
Record number :
1514552
Link To Document :
بازگشت