Characterization of thermal desorption instrumentation with a direct liquid deposition calibration method for trace 2,4,6-trinitrotoluene quantitation
Author/Authors :
Field، نويسنده , , Christopher R. and Giordano، نويسنده , , Braden C. and Rogers، نويسنده , , Duane A. and Lubrano، نويسنده , , Adam L. and Rose-Pehrsson، نويسنده , , Susan L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
9
From page :
10
To page :
18
Abstract :
The use of thermal desorption systems for the analysis of trace vapors typically requires establishing a calibration curve from vapors generated with a permeation tube. The slow equilibration time of permeation tubes causes such an approach to become laborious when covering a wide dynamic range. Furthermore, many analytes of interest, such as explosives, are not available as permeation tubes. A method for easily and effectively establishing calibration curves for explosive vapor samples via direct deposition of standard solutions on thermal desorption tubes was investigated. The various components of the thermal desorption system were compared to a standard split/splitless inlet. Calibration curves using the direct liquid deposition method with a thermal desorption unit coupled to a cryo-focusing inlet were compared to a standard split/splitless inlet, and a statistical difference was observed but does not eliminate or deter the use of the direct liquid deposition method for obtaining quantitative results for explosive vapors.
Keywords :
4 , quantitation , 6-trinitrotoluene (TNT) , thermal desorption , Cryo-focusing , 2 , Explosives , GC-ECD
Journal title :
Journal of Chromatography A
Serial Year :
2012
Journal title :
Journal of Chromatography A
Record number :
1514922
Link To Document :
بازگشت