Ion-pair ultra-high performance liquid chromatographic analysis of monoamines: Peak-splitting at high flow rates
Author/Authors
Van Schoors، نويسنده , , Jolien and Brouwer، نويسنده , , Hendrik-Jan and Maes، نويسنده , , Katrien and Michotte، نويسنده , , Yvette and Van Eeckhaut، نويسنده , , Ann، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
7
From page
73
To page
79
Abstract
The use of ion-pair ultra-high performance liquid chromatography (UHPLC) coupled with electrochemical detection (ECD) is of great interest for the fast and sensitive determination of the monoamine neurotransmitters dopamine, noradrenaline and serotonin in microdialysis samples. However, when applying high flow rates in ion-pair UHPLC, other peaks than the initial compound peaks appear on the chromatogram. This peak-splitting phenomenon is caused by disturbed ion-pair retention mechanisms. The influence of several chromatographic parameters is investigated. Peak-splitting is delayed to higher flow rates when increasing the concentration of ion-pair reagent or buffering agent in the mobile phase, when decreasing the percentage of organic modifier in the mobile phase, when applying a stationary phase with a smaller amount of packing material or when increasing the separation temperature. One or a combination of these conditions can be applied to analyze the monoamine neurotransmitters using ion-pair UHPLC–ECD at high flow rates.
Keywords
UHPLC , Ion-pair chromatography , flow rate , Peak-splitting , Monoamines , Electrochemical detection
Journal title
Journal of Chromatography A
Serial Year
2013
Journal title
Journal of Chromatography A
Record number
1515442
Link To Document