Title of article :
Membrane-assisted solvent extraction of triazines and other semi-volatile contaminants directly coupled to large-volume injection–gas chromatography–mass spectrometric detection
Author/Authors :
Hauser، نويسنده , , Barbara and Popp، نويسنده , , Peter and Kleine-Benne، نويسنده , , Eike، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
10
From page :
27
To page :
36
Abstract :
A simple device was developed for in-vial liquid–liquid extraction using a polymer membrane (nonporous polypropylene) to separate an aqueous sample from an organic extractant. The membrane consisted of tubing with an internal diameter of 6 mm and a wall thickness of 0.05 mm, which was heat-sealed at the lower end and filled with 500 μl hexane. This membrane bag was incorporated into a conventional 20 ml headspace vial suitable for a multi-purpose sampler (MPS 2, Gerstel, Mülheim, Germany) directly interfaced to a gas chromatograph with a mass-selective detector. The sampler enabled the extraction vial to be mixed at a defined temperature with subsequent large-volume injection of the organic extract taken from the membrane bag. The method was evaluated using several triazines, 2,4-dichloroaniline, α-hexachlorocyclohexane and phenanthrene as model compounds. Extraction parameters such as temperature, agitation speed, and extraction time were optimised. Recoveries of 60–90% were achieved after 30 min extraction. By increasing the injection volume to 100 μl, detection limits of 1–10 ng/l were determined.
Keywords :
triazines
Journal title :
Journal of Chromatography A
Serial Year :
2002
Journal title :
Journal of Chromatography A
Record number :
1515884
Link To Document :
بازگشت