Title of article :
Mass transfer mechanism in chiral reversed phase liquid chromatography
Author/Authors :
Gritti، نويسنده , , Fabrice and Guiochon، نويسنده , , Georges، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
The mechanism of mass transfer in chiral chromatography was investigated using an experimental protocol already applied in RPLC and HILIC chromatography. The different contributions to the reduced height equivalent to a theoretical plate (HETP) include the longitudinal diffusion HETP term, the solid–liquid mass transfer resistance HETP term, the short-range eddy dispersion HETP term, and the long-range eddy dispersion HETP term. Their accurate measurement permits the determination of the adsorption rate constant kads of trans-stilbene enantiomers on a column packed with Lux 5 μm Cellulose-1 particles. The experimental results demonstrate that the number of adsorption–desorption steps per unit time of chiral compounds on polysaccharide-based chiral stationary phases is four orders of magnitude smaller than that of achiral compounds.
Keywords :
Adsorption–desorption steps , Tri-tert-butylbenzene , ?m Cellulose-1 , trans-Stilbene , Chiral separation , Mass transfer resistance , Lux 5
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A