Title of article :
A new validated analytical method for the determination of tributyltin in water samples at the quantification level set by the European Union
Author/Authors :
Devos، نويسنده , , Christophe and David، نويسنده , , Frank and Sandra، نويسنده , , Pat، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Abstract :
According to recent directives of the European Union (EU), limits of quantification (LOQ) for the determination of tributyltin (TBT) in surface waters should be ca. 60 pg/L (ppq). This put very stringent requirements on analytical methodologies; definitely when they have to be applied in a routine environment. Stir bar sorptive extraction (SBSE), followed by thermal desorption (TD) and capillary gas chromatography–triple quadrupole mass spectrometry (GC–MS/MS) can provide accurate and precise data at the 2 ng/L level (ppt). For lower concentrations, matrix and reagent interferences together with contamination may provide too high TBT values. A two-dimensional heart-cut GC method was developed to fractionate TBT from interferences. The GC–GC–MS/MS method shows excellent linearity in the range 50 pg/L–4 ng/L, good repeatability (RSD < 20% at 200 pg/L), and a limit of detection of 11 pg/L. The method performance is demonstrated with representative samples i.e. harbor water and waste water samples.
Keywords :
Capillary gas chromatography–triple quadrupole mass spectrometry , Two-dimensional GC , Tributyltin , Stir bar sorptive extraction , Ultra-trace analysis
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A