Title of article :
Field enhanced bacterial sample stacking in isotachophoresis using wide-bore capillaries
Author/Authors :
Joseph and Oukacine، نويسنده , , Farid and Quirino، نويسنده , , Joselito P. and Destoumieux-Garzَn، نويسنده , , Delphine and Cottet، نويسنده , , Hervé، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
5
From page :
180
To page :
184
Abstract :
The isotachophoretic analysis of different bacterial strains was studied using capillaries with different internal diameters from 50 to 250 μm. Several injection modes were investigated and compared in order to improve the limit of detection of bacteria by capillary isotachophoresis. A system suitability test obtained from the separation voltage was developed to ensure reliable results. As expected, the use of wider bore capillaries improved the analytical sensitivity of the isotachophoretic method when compared to the 50 μm capillary. With the optimized conditions, the isotachophoretic method presented in this work allows the quantification of Erwinia carotovora (Gram negative bacteria) with a limit of detection as low as ∼3000 cells mL−1. The proposed methodology does not require any additive in the electrolyte such a fluorescent or chromophoric dye to reach these limits of detection.
Keywords :
Bacteria , Electrokinetic injection , Field-enhanced sample injection , Capillary zone electrophoresis , Isotachophoresis