Title of article :
Determination of carnitine and acylcarnitines in urine by high-performance liquid chromatography–electrospray ionization ion trap tandem mass spectrometry
Author/Authors :
Vernez، نويسنده , , Laurence and Hopfgartner، نويسنده , , Gérard and Wenk، نويسنده , , Markus and Krنhenbühl، نويسنده , , Stephan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
11
From page :
203
To page :
213
Abstract :
A high-performance liquid chromatography–mass spectrometry method has been developed for the simultaneous determination of native carnitine and eight acylcarnitines in urine. The procedure uses a solid-phase extraction on a cation-exchange column and the separation is performed without derivatization within 17 min on a reversed-phase C8 column in the presence of a volatile ion-pairing reagent. The detector was an ion trap mass spectrometer and quantification was carried out in the MS–MS mode. Validation was done for aqueous standards at ranges between 0.75 and 200 μmol/l, depending on the compound. Carnitine was quantified in urine and comparison with a radioenzymatic assay gave a satisfactory correlation (R2=0.981). The assay could be successfully applied to the diagnostic of pathological acylcarnitines profile of metabolic disorders in urines of patients suffering from different organic acidurias.