Title of article :
Explanation for the enhanced dissolution of silica column packing in high pH phosphate and carbonate buffers
Author/Authors :
Tindall، نويسنده , , G.W. and Perry، نويسنده , , R.L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
It has been reported that at high pH, the rate of bonded phase packing degradation in methanol/water mobile phases is greater for carbonate and phosphate buffers than for amine buffers. This conclusion was based on buffer pH determined in the aqueous buffer before dilution with methanol. Changes in buffer species pKa, and therefore buffer pH, upon methanol dilution are consistent with the observed degradation results. Measurements of pH in the methanol/water solutions confirm that the carbonate and phosphate buffers were considerably more basic than the amine buffer, even though all the buffers were pH 10 before dilution with methanol. These results demonstrate that it can be misleading to extrapolate aqueous pH data to partially aqueous solutions. Measurements of pH in the mixed solvent provide more reliable predictions of column and sample stability.
Journal title :
Journal of Chromatography A
Journal title :