Title of article
Enhanced mass resolution tandem mass spectrometry method for 2,3,7,8-tetrachlorodibenzo-p-dioxin detection with ion trap mass spectrometry using high damping gas pressure
Author/Authors
Kemmochi، نويسنده , , Yukio and Tsutsumi، نويسنده , , Kaori and Nakazawa، نويسنده , , Hiroyuki، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
8
From page
249
To page
256
Abstract
Damping gas flow was optimized for 2,3,7,8-tetrachlorodibenzo-p-dioxin (TCDD) determination using ion trap mass spectrometer. A tandem mass spectrometry (MS–MS) method with better than unit-mass resolution (mass width, 0.3 u) was developed at a damping gas flow of 1.5 ml/min and a collision-induced dissociation (CID) voltage of 3.30 V. The relative standard deviation (R.S.D.) at the enhanced resolution was 2.9% in 24 h of consecutive injections. The detection limit was significantly improved because the efficiency of both precursor ion trapping and fragmentation increased with the damping gas flow. Product ion yield was 4.5 times higher and limit of detection was 3.2 times lower than at the default flow (0.3 ml/min and 1.65 V).
Keywords
Damping gas flow , mass spectrometry , Polychlorinated Dibenzodioxins
Journal title
Journal of Chromatography A
Serial Year
2003
Journal title
Journal of Chromatography A
Record number
1517928
Link To Document