Title of article :
Screening of inorganic gases released from firework-rockets by a gas chromatography/whistle-accelerometer method
Author/Authors :
Chen، نويسنده , , Kuan-Fu and Wu، نويسنده , , Hui-Hsin and Lin، نويسنده , , Chien-Hung and Lin، نويسنده , , Cheng-Huang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
4
From page :
230
To page :
233
Abstract :
The use of an accelerometer for detecting inorganic gases in gas chromatography (GC) is described. A milli-whistle was connected to the outlet of the GC capillary and was used instead of a classical GC detector. When the GC carrier gases and the sample gases pass through the milli-whistle, a sound is produced, leading to vibrational changes, which can be recorded using an accelerometer. Inorganic gases, including SO2, N2 and CO2, which are released from traditional Chinese firework-rockets at relatively high levels as the result of burning the propellant and explosive material inside could be rapidly determined using the GC/whistle-accelerometer system. The method described herein is safe, the instrumentation is compact and has potential to be modified so as to be portable for use in the field. It also can be used in conjunction with FID (flame ionization detector) or TCD (thermal conductivity detector), in which either no response for FID (CO2, N2, NO2, SO2, etc.) or helium gas is needed for TCD, respectively.
Keywords :
Whistle , Gas analysis of firework-rockets , Accelerometer , Gas chromatography
Journal title :
Journal of Chromatography A
Serial Year :
2013
Journal title :
Journal of Chromatography A
Record number :
1518005