Title of article :
Sample pre-concentration by isotachophoresis in microfluidic devices
Author/Authors :
Wainright، نويسنده , , Ann and Williams، نويسنده , , Stephen J and Ciambrone، نويسنده , , Gary and Xue، نويسنده , , Qifeng and Wei، نويسنده , , Jing and Harris، نويسنده , , Dennis، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
12
From page :
69
To page :
80
Abstract :
We have designed microfluidic devices with the aim of coupling isotachophoresis (ITP) with zone electrophoresis (ZE) as a method to increase the concentration limit of detection in microfluidic devices. We used plastic multi-channel chips, designed with long sample injection channel segments, to increase the sample loading. The chip was designed to allow stacking of the sample into a narrow band by discontinuous ITP buffers and subsequent separation in the ZE mode. In the ITP–ZE mode, with a 2-cm long sample injection plug, sensitivity was increased by 400-fold over chip ZE and we found that the separation performance after the ITP stacking was comparable to that of regular chip ZE. We report sub-picomolar limits of detection of fluorescently labeled ACLARA eTag reporter molecules electrokinetically injected from cell lysate sample matrixes containing moderate salt concentrations. We evaluated sample injections from buffers with varied ionic strengths and found that efficient stacking and separations were obtained in both low and high conductivity buffers, including physiological buffer with at least 140 mM salt. We applied ITP–ZE to the analysis of a cell surface protease (ADAM 17) which used live intact cells in physiological buffers with detection limits below 10 cells/assay.
Keywords :
Peptides
Journal title :
Journal of Chromatography A
Serial Year :
2002
Journal title :
Journal of Chromatography A
Record number :
1518728
Link To Document :
بازگشت