Title of article :
Microcolumns with self-assembled particle frits for proteomics
Author/Authors :
Ishihama، نويسنده , , Yasushi and Rappsilber، نويسنده , , Juri and Andersen، نويسنده , , Jens S and Mann، نويسنده , , Matthias، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
7
From page :
233
To page :
239
Abstract :
LC–MS–MS experiments in proteomics are usually performed with packed microcolumns employing frits or outlets smaller than the particle diameter to retain the packing material. We have developed packed microcolumns using self-assembled particles (SAPs) as frits that are smaller than the size of the outlet. A five to one ratio of outlet size to particle diameter appears to be the upper maximum. In these situations the particles assembled into an arch over the outlet like the stones in a stone bridge. When 3 μm particles were packed into a tapered column with an 8 μm outlet, two particles bridged the outlet with 0.3 pl dead volume and perfect success rate. In peptide analysis by LC–MS, the peak width at half height was normally less than 6 s, compared to 12 s without SAPs. The LC–MS–MS system provided 37% sequence coverage (21 matched peptides) for a tryptically-digested sample of 10 fmol bovine serum albumin. We also describe application of the SAP principle to make disposable pipette tip columns with short pieces of fused-silica capillary as the outlet.
Keywords :
Proteins
Journal title :
Journal of Chromatography A
Serial Year :
2002
Journal title :
Journal of Chromatography A
Record number :
1518747
Link To Document :
بازگشت