Title of article :
Trace analysis of benzene, toluene, ethylbenzene and xylene isomers in environmental samples by low-pressure gas chromatography–ion trap mass spectrometry
Author/Authors :
Joos، نويسنده , , P.E. and Godoi، نويسنده , , A.F.L. and De Jong، نويسنده , , R. and de Zeeuw، نويسنده , , J. and Van Grieken، نويسنده , , R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
6
From page :
191
To page :
196
Abstract :
A rapid determination of benzene, toluene, ethylbenzene and the three xylene isomers (BTEX), including a nearly baseline separation of the xylene isomers in environmental samples within 1 min has been carried out using low-pressure gas chromatography–ion trap mass spectrometry (LP-GC–IT-MS). In order to evaluate the different parameters which may influence the performance of LP-GC–IT-MS, different column and mass spectral parameters were varied. Comparing LP-GC–IT-MS with the conventional equivalent, we obtained excellent detection limits as well as a good RSD of 8–13% in addition to a much shorter analysis time. In order to evaluate LP-GC–IT-MS for use in environmental samples, we determined BTEX in air.
Keywords :
Benzene , Ethylbenzene , Toluene , Xylenes
Journal title :
Journal of Chromatography A
Serial Year :