Title of article :
Separation of peptides by pressurized capillary electrochromatography
Author/Authors :
Zhang، نويسنده , , Kai and Jiang، نويسنده , , Zhengjin and Yao، نويسنده , , Chuanyi and Zhang، نويسنده , , Zhichao and Wang، نويسنده , , Qinsun and Gao، نويسنده , , Ruyu and Yan، نويسنده , , Chao، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
6
From page :
453
To page :
458
Abstract :
A pressurized electrochromatography (pCEC) instrument with gradient capability was used in this work for separation of peptides. Three separation modes, namely, pCEC, high-performance liquid chromatography and capillary electrophoresis can be carried out with the instrument. In pCEC mode, the mobile phase is driven by both electroosmotic flow and pressurized flow, facilitating fine-tuning in selectivity of neutral and charged species. A continuous gradient elution can be carried out conveniently on this instrument, which demonstrates that it is more powerful than isocratic pCEC for separation of complicated samples. The effects of applied voltage, supplementary pressure and ion-pairing agents on separation of peptides in gradient pCEC were investigated. The effects of flow-rate of the pump and the volume of the mixer on resolution were also evaluated.
Keywords :
Peptides
Journal title :
Journal of Chromatography A
Serial Year :
2003
Journal title :
Journal of Chromatography A
Record number :
1518926
Link To Document :
بازگشت