Title of article :
Integration of a contactless conductivity detector into a commercial capillary cassette: Detection of inorganic cations and catecholamines
Author/Authors :
Vuorinen، نويسنده , , Pasi S and Jussila، نويسنده , , Matti and Sirén، نويسنده , , Heli and Palonen، نويسنده , , Sami and Riekkola، نويسنده , , Marja-Liisa، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
8
From page :
45
To page :
52
Abstract :
A contactless conductivity detector integrated into the capillary cassette of Agilent 3DCE equipment is described. The detector is user-friendly, compact and easily modified. The UV detector of the 3DCE equipment is available parallel with the contactless conductivity detector increasing the detection power. Two electrolyte solutions, 2-(N-morpholino)ethanesulfonic acid–histidine solution (20 mM, pH 6.0) and ammonium acetate (10 mM, pH 4.0), were used as the separation media for inorganic cations and organic catecholamines, respectively. The detection limit for all metal cations except barium was under 0.5 mg/l, and that for four catecholamines was ca. 10 mg/l. This last value was the same order of magnitude as achieved with parallel UV detection.
Keywords :
Inorganic cations , catecholamines
Journal title :
Journal of Chromatography A
Serial Year :
2003
Journal title :
Journal of Chromatography A
Record number :
1518941
Link To Document :
بازگشت