• Title of article

    Comparative study on conventional and low-flow nebulizers for arsenic speciation by means of microbore liquid chromatography with inductively coupled plasma mass spectrometry

  • Author/Authors

    Sun، نويسنده , , Y.C. and LEE، نويسنده , , Y.S. and Shiah، نويسنده , , T.L. and Lee، نويسنده , , P.L. and Tseng، نويسنده , , W.C. and Yang، نويسنده , , M.H.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    207
  • To page
    213
  • Abstract
    The performance of conventional and low-flow nebulizer systems with liquid chromatography in differentiating four arsenic species in urine was evaluated. Two low-flow (DIN and MCN) chamber assemblies and a conventional (CFN) nebulizer-spray chamber assembly were compared in the hyphenation of anion-exchange microbore liquid chromatography with inductively coupled plasma mass spectrometry. Under optimal analytical conditions, the detection limits of the four arsenic species were 0.2–0.6 ng ml−1 for all the nebulizer systems tested. The chromatographic resolution was best in the case of DIN due to its minimal off-column dead volume and superior transport efficiency. Four arsenic species were determined in the certified reference materials NIST SRM 2670E and 2670N.
  • Keywords
    Arsenic
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography A
  • Record number

    1519387