• Title of article

    Multiple-detector responses or multiple-retention times: what is more informative for gas chromatography peak identification?

  • Author/Authors

    Stoev، نويسنده , , G. and Mihailova، نويسنده , , A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    7
  • From page
    263
  • To page
    269
  • Abstract
    Multiple-separation and -detection are two approaches applied at the identification of analytes in chromatography. Using them depends on the physico-chemical properties and elemental content of the analytes. When physico-chemical properties are similar multiple-separation gives better opportunities for the identification. In this case, the efficiency of the columns is very important. When analytes contain some characteristic groups as NO2, halogen, or nitrogen atoms then multiple-detection will be more useful. The sensitivity and/or selectivity of the detectors increase reliability of identification significantly.
  • Keywords
    detection , GC , Retention times , Selectivity , efficiency , Sensitivity
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography A
  • Record number

    1520210