• Title of article

    Evaluation of packed capillary liquid chromatography columns and comparison with conventional-size columns

  • Author/Authors

    S.M. and Prüك، نويسنده , , Anja and Kempter، نويسنده , , Christine and Gysler، نويسنده , , Jens and Jira، نويسنده , , Thomas، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    10
  • From page
    167
  • To page
    176
  • Abstract
    Apart from extracolumn effects peak dispersion in liquid chromatographic columns is caused by the column inlet, the packed bed, and the column outlet. A strategy applicable for independent evaluation of the individual sources of column band broadening was developed on the basis of the linear extrapolation method (LEM). This method was applied to compare the performance of packed capillary LC columns from various commercial suppliers with conventional-size columns. The columns differed widely in their performance with respect to peak shapes and widths for standard substances. The capillary columns were found well packed, but in some cases overall performance would benefit from improving the design of the area between the packed bed and the connecting capillaries, containing frits as well as dead volumes.
  • Keywords
    Stationary phases , LC , Packed capillary columns , Column performance , Alkylbenzenes , uracil , Toluene , linear extrapolation method
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography A
  • Record number