Title of article :
Challenges encountered in extending the sensitivity of US Environmental Protection Agency Method 314.0 for perchlorate in drinking water
Author/Authors :
Wagner، نويسنده , , Herbert P and Suarez، نويسنده , , F.Xavier and Pepich، نويسنده , , Barry V. and Hautman، نويسنده , , Daniel P. and Munch، نويسنده , , David J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
8
From page :
97
To page :
104
Abstract :
Concerns about the potential adverse health effects of perchlorate at concentrations below the minimum reporting level (MRL) of US Environmental Protection Agency (EPA) Method 314.0 (generally recognized as 4.0 μg/l) have led to an interest in increasing the sensitivity of the method. This work describes the use of 2 mm columns with a large-loop direct injection method, a column concentration technique and this concentration technique with a background reduction step, to increase the sensitivity for the analysis of trace levels of perchlorate in high ionic strength matrices. The concentrator columns studied were the Dionex TAC LP-1 and a new Dionex high capacity Cryptand concentrator column. The use of a surrogate to monitor trapping efficiency for the concentration technique and the use of confirmational columns to minimize the potential for false positives are also discussed. The large-loop direct injection method and the column concentration methods provided acceptable data when the samples were pre-treated with solid phase pretreatment cartridges. The background reduction technique did not provide acceptable data with either of the concentrator columns evaluated.
Keywords :
Water analysis , Perchlorate
Journal title :
Journal of Chromatography A
Serial Year :
2004
Journal title :
Journal of Chromatography A
Record number :
1520522
Link To Document :
بازگشت