Title of article
Intermittent counter-current extraction—Equilibrium cell model, scaling and an improved bobbin design
Author/Authors
Hewitson، نويسنده , , Peter E. Sutherland، نويسنده , , Ian and Kostanyan، نويسنده , , Artak E. and Voshkin، نويسنده , , Andrei A. and Ignatova، نويسنده , , Svetlana، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
10
From page
18
To page
27
Abstract
This paper describes an equilibrium cell model for intermittent counter-current extraction that is analytically solved for the first time for continuous sample injection between a pair of columns. The model is compared with practice for injections of a model mixture of compounds on a standard high-performance counter-current chromatography instrument giving good agreement for compound elution order and the times to maximum concentration for the eluted components. An improved design of end fittings for the counter-current chromatography bobbins is described which permits on-column switching of the mobile and stationary phases. This on-column switching successfully eliminates the displaced stationary phase seen in fractions when operating ICcE with standard flying leads and gives a 6% reduction in the retention time of compounds and improved resolution due to the elimination of the time delay required to pump the previous mobile phase from standard flying leads.
Keywords
Intermittent counter-current extraction , Flying leads , Equilibrium cell model , scale up , ICcE