Title of article
Estimation of the uncertainty associated with the results based on the validation of chromatographic analysis procedures: Application to the determination of chlorides by high performance liquid chromatography and of fatty acids by high resolution gas chr
Author/Authors
Quintela، نويسنده , , M. and Bلguena، نويسنده , , J. and Gotor، نويسنده , , G. A. Blanco Heras، نويسنده , , M.J. and Broto، نويسنده , , F.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
11
From page
107
To page
117
Abstract
This article presents a model to calculate the uncertainty associated with an analytical result based on the validation of the analysis procedure. This calculation model is proposed as an alternative to commonly used bottom-up and top-down methods. This proposal is very advantageous as the validation of the procedures and the estimation of the uncertainty of the measurement are part of the technical requirements needed in order to obtain the ISO 17025:2005 accreditation. This model has been applied to the determination of chloride by liquid chromatography in lixiviates and in the determination of palmitic acid and stearic acid by gas chromatography in magnesium stearate samples.
Keywords
Validation , uncertainty , Gas chromatography , Liquid chromatography
Journal title
Journal of Chromatography A
Serial Year
2012
Journal title
Journal of Chromatography A
Record number
1521034
Link To Document