Title of article :
Trace anion determination in concentrated hydrofluoric acid solutions by two-dimensional ion chromatography: II. Method performance study with a hydroxide eluent and a low noise suppressor
Author/Authors :
Vermeiren، نويسنده , , K.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
8
From page :
66
To page :
73
Abstract :
The two-dimensional ion exclusion chromatography/ion chromatography (ICE–IC) approach is considered to be the method of choice for the determination of trace anions in concentrated hydrofluoric acid (48–50%, w/w). In order to achieve lower detection limits, this method was for the first time used with electrolytically generated and purified hydroxide eluents in combination with a low noise electrochemical suppressor. Compared to carbonate based eluents, the achieved gain in peak height sensitivity for chloride, sulfate, nitrate and phosphate is a factor of 6, 7, 16 and 13, respectively. The instrumental detection limits, based on the background noise, are 2, 0.2, 0.4 and 1.4 μg/kg HF 50% (w/w) for the same anions. Their method detection limits, calculated according to SEMI, are all within the 6–10 μg/kg HF 50% (w/w) range and thus at least 10 times lower than the current Tier C grade requirements. The chromatographic run time could be shortened with some 10 min by the use of a relatively fast high-capacity hydroxide selective anion exchange column.
Keywords :
trace analysis , Two-dimensional ion chromatography , SEMI , Method validation , Hydrofluoric acid
Journal title :
Journal of Chromatography A
Serial Year :
2005
Journal title :
Journal of Chromatography A
Record number :
1521202
Link To Document :
بازگشت