• Title of article

    Trace anion determination in concentrated hydrofluoric acid solutions by two-dimensional ion chromatography: II. Method performance study with a hydroxide eluent and a low noise suppressor

  • Author/Authors

    Vermeiren، نويسنده , , K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    8
  • From page
    66
  • To page
    73
  • Abstract
    The two-dimensional ion exclusion chromatography/ion chromatography (ICE–IC) approach is considered to be the method of choice for the determination of trace anions in concentrated hydrofluoric acid (48–50%, w/w). In order to achieve lower detection limits, this method was for the first time used with electrolytically generated and purified hydroxide eluents in combination with a low noise electrochemical suppressor. Compared to carbonate based eluents, the achieved gain in peak height sensitivity for chloride, sulfate, nitrate and phosphate is a factor of 6, 7, 16 and 13, respectively. The instrumental detection limits, based on the background noise, are 2, 0.2, 0.4 and 1.4 μg/kg HF 50% (w/w) for the same anions. Their method detection limits, calculated according to SEMI, are all within the 6–10 μg/kg HF 50% (w/w) range and thus at least 10 times lower than the current Tier C grade requirements. The chromatographic run time could be shortened with some 10 min by the use of a relatively fast high-capacity hydroxide selective anion exchange column.
  • Keywords
    trace analysis , Two-dimensional ion chromatography , SEMI , Method validation , Hydrofluoric acid