• Title of article

    Determination of l-carnitine in food supplement formulations using ion-pair chromatography with indirect conductimetric detection

  • Author/Authors

    Kakou، نويسنده , , Aikaterini and Megoulas، نويسنده , , Nikolaos C. and Koupparis، نويسنده , , Michael A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    209
  • To page
    215
  • Abstract
    A novel method for the determination of l-carnitine in food supplement formulations was developed and validated, using ion-pair chromatography with indirect conductimetric detection. The chromatographic method was based on a non-polar (C18) column and an aqueous octanesulfonate (0.64 mM) eluent, acidified with trifluoroacetic acid (5.2 mM). The retention time was 5.4 min and the asymmetry factor 0.65. A linear calibration curve from 10 to 1000 μg/ml (r = 0.99998), with a detection limit of 2.7 μg/ml (25 μl injection volume), a repeatability %RSD of 0.8 (40 μg/ml, n = 5) and reproducibility %RSD of 2.6 were achieved. The proposed method was applied for the determination of carnitine in oral solutions and capsules. No interference from excipients was found and the only pretreatment step required was the appropriate dilution with the mobile phase. Recovery from spiked samples was ranged from 97.7 to 99.7% with a precision (%RSD, n = 3) of 0.01–2.1%.
  • Keywords
    L-carnitine , Levocarnitine , Ion-pair chromatography , Food supplement , Indirect conductimetric detection
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2005
  • Journal title
    Journal of Chromatography A
  • Record number

    1521318