• Title of article

    Development and validation of a hydrophilic interaction chromatography-mass spectrometry assay for taurine and methionine in matrices rich in carbohydrates

  • Author/Authors

    Marine de Person، نويسنده , , Marine and Hazotte، نويسنده , , Aurélie and Elfakir، نويسنده , , Claire and Lafosse، نويسنده , , Michel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    8
  • From page
    174
  • To page
    181
  • Abstract
    A new procedure based on hydrophilic interaction chromatography coupled to tandem mass spectrometry (ionisation process by pneumatically assisted electrospray in negative ion mode), is developed and validated for the simultaneous determination of underivatised taurine and methionine in beverages rich in carbohydrates such as energy drinks. No initial clean-up procedure and no sample derivatisation are required. Satisfactory analysis was obtained on an Astec apHera NH2 (150 mm × 4.6 mm; 5 μm) column with methanol–water (60/40) as mobile phase. The method was validated in terms of specificity, detection limits, linearity, accuracy, precision and stability, using threonine as internal standard. The potential effects of matrix and endogenous amino acid content were also examined. The limits of detection in the beverage varied from 20 μg L−1 for taurine to 50 μg L−1 for methionine.
  • Keywords
    Methionine , Assay validation , Underivatised amino acids , Hydrophilic interaction chromatography , mass spectrometry , Taurine
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2005
  • Journal title
    Journal of Chromatography A
  • Record number

    1521402