Abstract :
A new procedure based on hydrophilic interaction chromatography coupled to tandem mass spectrometry (ionisation process by pneumatically assisted electrospray in negative ion mode), is developed and validated for the simultaneous determination of underivatised taurine and methionine in beverages rich in carbohydrates such as energy drinks. No initial clean-up procedure and no sample derivatisation are required. Satisfactory analysis was obtained on an Astec apHera NH2 (150 mm × 4.6 mm; 5 μm) column with methanol–water (60/40) as mobile phase. The method was validated in terms of specificity, detection limits, linearity, accuracy, precision and stability, using threonine as internal standard. The potential effects of matrix and endogenous amino acid content were also examined. The limits of detection in the beverage varied from 20 μg L−1 for taurine to 50 μg L−1 for methionine.
Keywords :
Methionine , Assay validation , Underivatised amino acids , Hydrophilic interaction chromatography , mass spectrometry , Taurine