Title of article :
Counter-current chromatography separation scaled up from an analytical column to a production column
Author/Authors :
Wood، نويسنده , , Philip and Ignatova، نويسنده , , Svetlana and Janaway، نويسنده , , Lee and Keay، نويسنده , , David and Hawes، نويسنده , , David and Garrard، نويسنده , , Ian W. Sutherland، نويسنده , , Ian A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
An analytical separation was performed on an analytical J-type counter-current chromatography (CCC) instrument using a 5.4 ml column, with a 1 ml/min mobile phase flow rate. This separation had a resolution of 0.69 and was achieved in 10 min. The same separation was performed using two 2300 ml columns connected in series at a flow rate of 850 ml/min using a production scale J-type centrifuge. This production scale separation was also obtained in 10 min with a resolution of 0.71. This represents an 850 times increase in productivity. This paper presents these separations and the underlying scale up theory.
Keywords :
scale up , High-performance liquid chromatography , High-performance counter-current chromatography , Mass transfer , Counter-current chromatography
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A