• Title of article

    Development of septum-free injector for gas chromatography and its application to the samples with a high boiling point

  • Author/Authors

    Ito، نويسنده , , Hiroshi and Hayakawa، نويسنده , , Kazuichi and Yamamoto، نويسنده , , Atsushi and Murase، نويسنده , , Atsushi and Hayakawa، نويسنده , , Kazumi and Kuno، نويسنده , , Minoru and Inoue، نويسنده , , Yoshinori، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    4
  • From page
    256
  • To page
    259
  • Abstract
    A novel apparatus with a simple structure has been developed for introducing samples into the vaporizing chamber of a gas chromatograph. It requires no septum due to the gas sealing structure over the carrier gas supply line. The septum-free injector made it possible to use injection port temperatures as high as 450 °C. Repetitive injection of samples with boiling points below 300 °C resulted in peak areas with relative standard deviations between 1.25 and 3.28% (n = 5) and good linearity (r2 > 0.9942) for the calibration curve. In the analysis of polycyclic aromatic hydrocarbons and a base oil, the peak areas of components with high boiling points increased as the injection port temperature was increased to 450 °C.
  • Keywords
    Sealing gas line , high temperature , Gas-chromatograph , Septum-free injector