Title of article :
Effect of residues remaining in the injection liner of a gas chromatograph on the quantification of polycyclic aromatic hydrocarbons by isotope dilution mass spectrometry using deuterium-labeled internal standards
Author/Authors :
Itoh، نويسنده , , Nobuyasu and Numata، نويسنده , , Masahiko and Aoyagi، نويسنده , , Yoshie and Yarita، نويسنده , , Takashi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
7
From page :
246
To page :
252
Abstract :
We have investigated the differences in response ratios of native polycyclic aromatic hydrocarbons (native PAHs)/13C-labeled PAHs (13C-PAHs) and of native PAHs/deuterium-labeled PAHs (PAHs-d) in a calibration solution containing at trace concentrations (<1 μg mL−1) by gas chromatography–mass spectrometry. Only the ratios of native PAHs/PAHs-d decreased significantly (p < 0.05) accompanied with increase of residue amounts originated from immediately before injected solution. Since the analytical results using PAHs-d as internal standards were significantly lower than those using 13C-PAHs whatever solutions were injected immediately before (p < 0.05, 1.9–13.2%), 13C-PAHs gave more accurate results.
Keywords :
Isotope dilution mass spectrometry (IDMS) , 13C-labeled , Polycyclic Aromatic Hydrocarbons (PAHs) , GC–MS , Matrix effect , Deuterium-labeled
Journal title :
Journal of Chromatography A
Serial Year :
2006
Journal title :
Journal of Chromatography A
Record number :
1522073
Link To Document :
بازگشت