Detection of dideuterium in diprotium gas at infinitesimal concentration using a trace reduction detector
Author/Authors
Kawano، نويسنده , , Takao and Tsuboi، نويسنده , , Naohiro and Tsujii، نويسنده , , Hirotsugu and Uda، نويسنده , , Tatsuhiko، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
6
From page
408
To page
413
Abstract
An analyzer able to detect extremely small concentrations of dideuterium in diprotium gas was developed. To examine its performance, four gas samples of pure diprotium with dideuterium concentrations of 0.25, 0.5, 1, and 5 cm3/m3 were measured using the analyzer, and peak areas and retention times from 18 chromatograms were determined. Results indicated that the peaks were distributed within a relative standard deviation of approximately 15% and a linear correlation existed between peak area and concentration. Retention times were stable, showing a relative standard deviation of 1% with repeated tests and having no correlation with concentration of the four gas samples.
Keywords
gas chromatograph , Atomic absorption , Infinitesimal dideuterium detection , Trace reduction detector , Pre- and post-cut methods