Title of article :
Size-exclusion chromatography using deuterated mobile phases
Author/Authors :
Erdner، نويسنده , , Jeanine M. and Barth، نويسنده , , Howard G. and Foley، نويسنده , , Joe P. and Payne، نويسنده , , William G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
The effect of deuterated solvents in size-exclusion chromatography (SEC) was studied by comparing intrinsic viscosity measurements, SEC calibration curves, and column efficiency using water-soluble polymers. For aqueous SEC, the use of deuterium oxide slightly increases the SEC elution volume. To verify that adsorption onto the packing was absent, data from exclusion experiments were compared at 35 and 50 °C. Our results indicate that adsorption is not occurring for pullulan or polyethylene glycol (PEG)/poly(ethylene oxide) (PEO); for the latter, however, the elution volume increased using both D2O and H2O, indicative of slight hydrodynamic volume contraction of PEG/PEO at higher temperatures. A moderate increase in band broadening (moderate decrease in column efficiency) was observed using D2O. Finally, the effects of chloroform versus deuterated chloroform were evaluated, but no hydrodynamic volume changes were observed.
Keywords :
Deuterated solvents , Deuterium oxide , Pullulan , PEG/PEO , Proteins , Gel filtration chromatography , gel permeation chromatography , SEC , size-exclusion chromatography
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A