• Title of article

    Detection enhancement in nano-channels using micro-machined silicon groove

  • Author/Authors

    Fekete، نويسنده , , Veronika and Clicq، نويسنده , , David and De Malsche، نويسنده , , Wim and Gardeniers، نويسنده , , Han and Desmet، نويسنده , , Gert، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    7
  • From page
    151
  • To page
    157
  • Abstract
    The present paper reports on an experimental study of the possibility to use a micro-machined detection groove to enhance the detection sensitivity in flat-rectangular nano-channels for ultra-rapid liquid chromatography separations. Transversally running detection grooves with three different axial widths (respectively, 2, 4 and 6 μm) and one depth (4.75 μm) were tested in glass and silicon channels for the whole range of detectable fluorescein isothiocyanate isomer I, FITC, concentrations. The groove with the most square-like cross-section (i.e., 4 μm wide and 4.75 μm deep) yielded the best combination of detection gain and minimal additional band broadening. In a 1 cm long channel, the effective plate loss caused by the 4 μm wide groove would only be of the order of 20%, while the gain in S/N-ratio was of the order of a factor of 5. The detection groove concept yields larger gains in silicon channel substrates than in glass channel substrates, due to the larger stray light losses occurring in the latter.
  • Keywords
    nanofluidics , Detection sensitivity , Silicon groove , Shear-driven chromatography , FITC
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2006
  • Journal title
    Journal of Chromatography A
  • Record number