Title of article :
Use of new generation poly(styrene-divinylbenzene) resins for gas-phase trapping-thermal desorption: Application to the retention of seven volatile organic compounds
Author/Authors :
Lَpez، نويسنده , , Patricia and Batlle، نويسنده , , Ramon and Nerيn، نويسنده , , Ana Cristina and Cacho، نويسنده , , Juan and Ferreira، نويسنده , , Vicente، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Two new generation polymeric resins, Bond Elut ENV (styrene-divinylbenzene) from Varian and LiChrolut EN (ethylvinylbenzene-divinylbenzene) from Merck, commonly used in liquid–solid-phase extraction (SPE) were evaluated as sorbents for gas-phase sampling followed by thermal desorption and compared to Tenax TA, a reference sorbent in this kind of applications. The three resins were tested against seven volatile organic compounds (VOCs): 1-octene, ethylbenzene, (p-, m-, o-)xylenes, styrene and 1,4-dichlorobenzene. Elution curves for all compounds were determined at temperatures from 120 to 180 °C, and from such curves, different parameters, such as retention factor (k), distribution coefficient (K), height equivalent to a theoretical plate (H), asymmetry factor (Fa) and breakthrough volume (VB) were calculated and extrapolated at room (25 °C) and desorption (220 °C) temperatures in order to estimate breakthrough and elution volumes. In average, retention in LiChrolut EN is 10 and 200 times stronger than in Bond Elut ENV and Tenax TA, respectively, but its chromatographic behavior is rather poor giving quite asymmetric elution profiles (Fa > 1.8 at 120 °C). Bond Elut ENV exhibited the best chromatographic behavior, with H values two or five times lower than those of LiChrolut EN or Tenax TA. An additional advantage of the new sorbents is that retention decreases with T much faster than it does in Tenax (8 or 20 times for Bond Elut ENV or LiChrolut EN). Modeling has finally shown that beds with 60–80 (for Bond Elut ENV) or 300–400 (for LiChrolut EN) times less of sorbent have the same retention properties than standard Tenax TA tubes and similar (LiChrolut EN) or five to six times smaller (Bond Elut ENV) elution volumes. These predictions have been experimentally confirmed.
Keywords :
Bond Elut ENV , LiChrolut EN , Solid-phase extraction , Distribution coefficients , Sorbent , Breakthrough volume , thermal desorption , Gas-phase sampling , Purge and trap , volatile organic compounds
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A